Revizie analizor hemoglobina glicozilata Arkray ADAMS HA-8180V

DA27928959 · published 2021-05-10 · completed 2021-05-10 · from Tehno Industrial (Bucuresti)

2 leiawarded value
—estimated value
50421000-2CPV code
The quantity is not published in the open data. The MS/SEAP export has only the title and value; quantity and unit price appear only in the purchase's SICAP record, which you can look up on e-licitatie.ro.

Description from SEAP

Revizia tehnica anuala include: - curatare si calibrare sistem optic cu materiale Arkray - curatare si calibrare sistem pneumatic - curatare, reglare si gresare sistem pipetare - curatare si gresare sistem mecanic transport rastele - curatare si verificare printer - curatare container de spalare - verificare functionare (cu consumabile si controale) + reglaje transportor rastele - curatare, calibrare, gresare si reglaje ansamblu Sampler - inlocuire filtre, o-ring-uri si tubulaturi uzate - eventuale reglaje ce vor fi constatate si declarate necesare de catre echipa noastra tehnica - raspunsuri la solicitari/cereri de remedieri ale unor defectiuni reclamate - inlocuirea coloanei si efectuarea calibrarii aferente noii coloane