Verificare MasterScreen Body-Diff

DA34076276 · published 2023-09-22 · completed 2023-09-25 · from Newmedics Com S.R.L. (Bucuresti)

1,150 leiawarded value
1,150 leiestimated value
98300000-6CPV code
The quantity is not published in the open data. The MS/SEAP export has only the title and value; quantity and unit price appear only in the purchase's SICAP record, which you can look up on e-licitatie.ro.

Description from SEAP

Verificare aparat Jaeger MasterScreen Body-Diff. 
Se verifica:
- ambient module
- pneumotach O-rings
- zero and gain at the PT-pressure transducer (use EASI-Test and 
reference PT-screen)
- zero (and gain*) at the PV-pressure transducer (use EASI-Test and 
reference pressure*)
- shutter valves for easy rotation (torque)
- shutter valve for leak
- PT-flange, fittings and plastic elbows for cracks (replace if necessary)
- zero and gain at PB- and PV-pressure transducer (use EASI-TEST and 
Body Calibration)
- flexible tube for wear and leak (replace if necessary)
- check temperature insulation at Helium sensor (p/n 845 568)
- analyzer (sample) pump alignment and adjust if necessary
- zero and gain factor at the CO analyzer fuel cell (use EASI-Test and gas 
calibration; analyzer fuel cells with gain factor >1.6 should be replaced)
- support arm tension (brake screw)
- 50ml sine pump
- Body Close Valve hub
- Pressure Safety Valve (if available) for leak and proper function (o-ring 
851032)
- all 3 door lock brackets (replace if necessary)
- body box for leak (use body box calibration)
Costul componentelor care necesita schimbare si manopera nu este inclus.
Nu toate componentele de service pot fi mentinute in permanenta in stocul distribuitorului;
componente de service care nu fac parte din procedura de intretinere preventiva si nu se afla in stoc vor fi comandate la producator, in cazul in care este necesara schimbarea lor; timpul standard de livrare este de 2-4 saptamani 
valabil