Reparatie Electrocauter OLYMPUS ESG-400, SN:B000492

DA41018106 · published 2026-08-19 · completed 2026-08-20 · from Tehno Electro Medical Company S.R.L. (Bucuresti)

45 k leiawarded value
45 k leiestimated value
50421000-2CPV code
The quantity is not published in the open data. The MS/SEAP export has only the title and value; quantity and unit price appear only in the purchase's SICAP record, which you can look up on e-licitatie.ro. The text does mention: 6 luni (read automatically from title/description; check the text below).

Description from SEAP

Reparatie Electrocauter OLYMPUS ESG-400, SN:B000492 Constatari: - la pornire, dispozitivul afiseaza eroarea E460 si se restarteaza; - eroarea este cauzata de defectiunea placii de baza si a sursei de alimentare de inalta tensiune (placa HVPS). - au fost identificate deteriorari si fisuri ale panoului frontal; Pentru a restabili functionarea corecta a dispozitivului, este necesara inlocuirea placii HVPS, a placii de baza si a panoului frontal; Garantie : 6 luni